Operating instructionsMETTLER TOLEDOB-S line of balances•AB-S•PB-SMETTLER TOLEDOFCal/MenuCOn offO/TMETTLER TOLEDOFCal/MenuCOn offO/TMETTLER
102.4 Adjusting (calibration)To obtain accurate weighing results, the balance must be adjusted to match the gravitational acceleration at itslocation.
113 Weighing3.1 On/off switchingSwitching on➞ Remove any load from weighing pan and press «On» key briefly.The balance performs a display test (all se
123.3 Taring➞ Place empty container on the balance.➞ The weight is displayed.➞ Tare: press the «#» key briefly.➞ Add weighing sample to container. The
134 Menu4.1 Overview and operationIn the menu you can change weighing units (only possible with certified balances if permitted by national weights an
14Using the menuOpening the menuIn weighing mode, press and hold down the «Cal/Menu» key until “MENU” appears in the display. Release thekey: the 1st
154.2 Description of menu options4.2.1 Reset or recording of balance settings (1st menu option “RESET”)Note: This menu option is only available if the
164.2.2 Weighing unit 1 (2nd menu option “UNIT 1”)Depending on requirements, the balance can operate with the following units:Unit Conversion factor C
174.2.4 Peripheral device (4th menu option)Note: This menu option is only available if the balance is equipped with the optional interface!At this men
184.2.7 Baud rate (7th menu option)Note: This menu option is only available if the balance is equipped with the optional interface and you haveselect
195.1 Piece counting➞ Place empty container on the balance and tare by briefly pressing the «#» key.Setting the reference: a reference weight must fir
Operating instructions in a nutshellPress key brieflyPress and hold keydown until the desireddisplay appearsautomatic sequencePiece countingSwitching
206 Technical data, interface, optional equipment6.1 Technical dataStandard equipment of B-S balances• Protective cover, transparent, made from Barex•
2121PB1502-S PB3002-S PB3002-S PB1501-S PB3001-S PB5001-S PB8001-STechnical Data DeltaRangeReadability 0.01 g 0.01 g 0.01 g*/0.1 g 0.1 g 0.1 g 0.1 g 0
22Auxiliary display• Auxiliary display for connecting to theoptional RS232C interface RS/LC-BLDAdjusting weightsAvailable as OIML weights (E1)(E2, F2,
23231)26619032110.5184.5Ø 1811089213245PB-S balance with no draft shield PB-S balance with low “mg” draft shield AB-S balance with high “0.1mg” draft
24Function: Adjusting- BALANCE CALIBRATION -Date: ...Time: ...METTLER TOLEDOType: PB3002-SSNR: 11180
25RectificationCause7.2 What if ...?Error/Error messageOverload➞ Remove sample from weighing pan, zero again (tare).Underload➞ Check whether weighing
267.3 Connecting B-S balances to other METTLER TOLEDO devicesDevice Connecting cable settings/RemarksTitrators:DL67, 70ES, 70 V.2, 77DL12, 18, 20, 21,
Printed on 100% chlorine-free paper, for the sake of our environment.
Subject to technical changes and to the availabilityof the accessories supplied with the instruments.To give your METTLER TOLEDO product an assured fu
3Contents1 Getting to know B-S balances ... 41.1 Introducing the B-S line of balances...
41.1 Introducing the B-S line of balancesSeveral types of construction – uniform operation• The B-S balance line ranges from high-resolution analytica
51.2 Layout of B-S balances1 Keys2 Display3 Model plate with the following data:“Max”:maximum capacity“d”: readability“Min”: minimum capacity (recomme
6--------------------Cal/Menu • Save changes and close menu1.3 Overview of key functionsThe balances have two operator control levels: the weighing mo
72 Startup2.1 Unpacking / standard equipmentAll B-S balances are supplied in environmentally compatible packaging. The standard equipment for everybal
82.3 Setting up, leveling, preparations for weighing below the balance, connecting to power supplyThe optimum locationThe correct location makes an im
9LevelingB-S balances have a spirit level and two adjustable leveling feet to compensate for slight irregularities in thesurface of the weighing bench