METTLER TOLEDO PS Shipping Scale Technical/Operators Manual
Error Code Section
Error codes are displayed on the weight indicator with a leading “E” to
distinguish themselves from weight data.
Code Meaning Action
E1 ROM (checksum) error Replace the circuit board to clear this error (call
Mettler Toledo Service)
E2 RAM error
Replace the circuit board to clear this error (call
Mettler Toledo Service)
E3 EEPROM error
Setup and /or calibration information has been lost.
Press any display button and the scale will enter setup
mode or calibration mode depending on what
information is lost.
If both setup and calibration information has been
lost, the scale will enter setup mode first when a
display button is pressed. After completing the setup,
the scale will display E3 again until a button is
pressed to enter calibration mode and calibration is
completed.
Reset the setup parameters and/or recalibrate to clear
this error. If the scale continues to display E3 after
the setup and/or calibration has been completed, the
load cell is unable to save the information and will
need to replaced (call Mettler Toledo Service)
E4 TC error
Replace the load cell to clear this error (call Mettler
Toledo Service)
E5 Serial Communications (Data) error
The scale has received a character from the host with
an invalid parity. The scale will display E5 until
either a display button is pressed or another character
is received. This error could be caused by an
incorrect Baud Rate or Data Bits selection in the scale
setup.
Wall Transformer
The wall transformer (when connected to the proper AC input and
disconnected from the scale) may read as high as 12.8 VDC. The wall
transformer should read from 9 VDC, up to 13 VDC when connected to the
scale. If the voltage measured is not within this range, replace the wall
transformer. The center conductor of the wall transformer is positive.
(08/04) 5-2