5.2 Model-specific data
Technical data (limit values)
Model XP155KS XP604KM XP1003KM XP2003KL XP6002KL
Maximum load 150 kg 600 kg 1100 kg 2500 kg 5400 kg
Readability 0.05 g 0.1 g 0.5 g 1 g 10 g
Repeatability at nominal load (ABA, measured at)
1)
0.12 g (100 kg) 0.23 g (500 kg) 1.5 g (1000 kg) 7 g (2000 kg) 70 g (5000 kg)
Repeatability at low load (ABA, measured at)
1)
0.09 g (5 kg) 0.15 g (20 kg) 1.0 g (50 kg) 4 g (100 kg) 50 g (500 kg)
Repeatability absolute (Full range) 0.15 g (100 kg) 0.3 g (500 kg) 2 g (1000 kg) 10 g (2000 kg) 100 g (5000 kg)
Electrical weighing range 0…150 kg 0…600 kg 0…1100 kg 0…2500 kg 0…5400 kg
Taring range 150 kg 600 kg 1100 kg 2500 kg 5400 kg
Pre-load with fullrange 0…125 kg 0…480 kg 0…100 kg 0…150 kg 0…440 kg
Linearity (electrical weighing range) 2.0 g (100 kg) 10 g (500 kg) 20 g (1000 kg) 100 g (2000 kg) 300 g (5000 kg)
Eccentric load deviation (measured at)
2)
5 g (50 kg) 40 g (200 kg) 40 g (200 kg) 120 g (1000 kg) 240 g (2000 kg)
Sensitivity offset 1x10
-5
·R
nt
1x10
-5
·R
nt
1x10
-5
·R
nt
1x10
-5
·R
nt
1 x10
-5
·R
nt
Sensitivity temperatur drift
3)
3x10
-6
/ºC·R
nt
3x10
-6
/ºC·R
nt
3x10
-6
/ºC·R
nt
3 x10
-6
/ºC·R
nt
3x10
-6
/ºC·R
nt
Sensitivity stability
4)
1.5 x10
-5
/a·R
nt
1.5 x10
-5
/a·R
nt
1.5 x10
-5
/a·R
nt
1.5x10
-5
/a·R
nt
1.5x10
-5
/a·R
nt
Settling time 5 s 5 s 5 s 5 s 5 s
Weighing time, 1 ABA measurement
5)
36…45 s 36…45 s 36…45 s 36…45 s 36…45 s
Interface update rate 23 /s 23 /s 23 /s 23 /s 23 /s
Internal adjustment weights 2 2 2 2 2
Adjustment built in proFACT proFACT proFACT proFACT proFACT
Measuring range of the external adjustment weights 50…150 kg 200…600 kg 200…1000 kg 500…2500 kg 1000…5000 kg
Weighing platform dimensions (W x D x H) [mm] 800 x 600 x 130 1000 x 800 x 115 1000 x 800 x 115 1500 x 1250 x 182 1500 x 1250 x 182
Terminal dimensions (W x D x H) [mm] 194 x 133 x 71 194 x 133 x 71 194 x 133 x 71 194 x 133 x 71 194 x 133 x 71
Weighing pan dimensions (W x D) [mm] 800 x 600 1000 x 800 1000 x 800 1500 x 1250 1500 x 1250
Weight [kg] 54 93 93 382 382
Typical data for determination of the measurement uncertainty
Model XP155KS XP604KM XP1003KM XP2003KL XP6002KL
Repeatability at nominal load (ABA, measured at)
1)
typical 0.09 g (100 kg) 0.18 g (500 kg) 1.0 g (1000 kg) 5.5 g (2000 kg) 60 g (5000 kg)
Repeatability at low load (ABA, measured at)
1)
typical 0.06 g (5 kg) 0.11 g (20 kg) 0.6 g (50 kg) 3.0 g (100 kg) 40 g (500 kg)
Repeatability (sd) typical 0.06 g+3.2x10-7·R
gr
0.11 g+6.3x10-8·R
gr
0.6 g+4.2x10-7·R
gr
3.0 g+1.3x10-6·R
gr
40 g+4.2x10-6·R
gr
Differential nonlinearity (sd) typical √ 5 x105 x10
-12
R
nt
√ 5 x105 x10
-12
R
nt
√ 5x105x10
-12
R
nt
√ 5 x10
-12
R
nt
√ 5 x105 x10
-12
R
nt
Differential eccentric load deviation (sd) typical 1.2 x10
-5
·R
nt
1.2x10
-5
·R
nt
1.2 x10
-5
·R
nt
3.3 x10
-5
·R
nt
4.0 x10
-5
·R
nt
Sensitivity offset (sd) typical
3)
2 x10
-6
·R
nt
2 x10
-6
·R
nt
2 x10
-6
·R
nt
3 x10
-6
·R
nt
3 x10
-6
·R
nt
Minimum weight (1%, 2sd) typical 12 g+ 4.2x10
-5
·R
gr
22 g+ 2.4x10
-5
·R
gr
120 g+ 7.6 x10
-5
·R
gr
600 g+ 2x10
-4
·R
gr
8000 g+ 7.8 x10
-4
·R
gr
R
gr
= gross weight
R
nt
= net weight (sample weight)
sd = Standard deviation
a = Year (annum)
1)
Value out of 5 ABA measurements according to OIML R111
2)
According to OIML R76
3)
In the temperature range 10…30 °C
4)
Sensitivity drift/year after putting into operation for the first time, with the ProFACT self-calibration function activated
5)
Weighing time without loading
Technical data
16
Page view 15
1 2 ... 11 12 13 14 15 16 17 18 19 20 21 ... 27 28

Comments to this Manuals

No comments