4.2 Description of the menu options
4.2.1 Reset and printout of the balance setting
Reset balance setting and functions to factory setting (rESEt)
Select "rESEt" and press and hold
Cal/Menu
until "r donE" is displayed.
The balance is now reset to the factory setting and returns to the weighing mode.
F nonE no function activated PrintEr Attachment to printer or host
_ 2 normal balance environment bd2400 Transmission rate
Unit 1 g 7b–E Character format
Unit 2 g HS oFF Transmission protocol
A. oFF – no automatic shutdown
Printing out the balance setting (LISt)
Select "LISt" and press and hold
Cal/Menu
until "StorEd" is displayed.
The current balance setting is printed out and stored.
4.2.2 F... – Selecting function of F key (detailed description in section 5)
In addition to simple weighing, the following functions can be selected:
F nonE No function, simple weighing (factory setting)
F count Piece counting
F 100 % Percent weighing
F dYn A Dynamic weighing with automatic start
F dYn M Dynamic weighing with manual start
4.2.3 Setting the vibration adapter
You can use the vibration adapter to match the balance to the ambient conditions.
_ 2 Setting with normal balance surroundings (factory setting).
3 Setting with unstable balance surroundings. The balance operates slower but is less sensitive to
external influences (drafts, vibrations, etc.).
1 Setting with very stable balance surroundings. The balance operates very quickly but is sensitive
to external influences (drafts, vibrations, etc.).
rESEt
LISt
F nonE
F dYn Ÿ≈
F count
F 100
%
F dYn A
_
2
3
1
_
2