Mettler Toledo XP26 Operations Instructions

Browse online or download Operations Instructions for Kitchen scales Mettler Toledo XP26. Mettler Toledo XP26 Operating instructions [en] User Manual

  • Download
  • Add to my manuals
  • Print

Summary of Contents

Page 1 - Operating Instructions

Operating InstructionsExcellence Plus BalancesXP Models – Part 3

Page 2

Settings for Adjustment and Tests103.2 Test SequencesTest Sequences define which test is to be performed and with which test weights(s). The user is g

Page 3 - Contents

Settings for Adjustment and Tests11When “Test Sequences” is selected you are provided with a list of tests. Up to 12 Test Sequences can be configured

Page 4 - 1.3 GWPExcellence™

Settings for Adjustment and Tests12“Preparation Instructions”: Choose between two settings: “None” 8: No preparation instructions will be included i

Page 5

Settings for Adjustment and Tests13“Instructions if Failure”: Defines instructions for the user that will be displayed after a test fails. This sett

Page 6

Settings for Adjustment and Tests14 “Advanced”: The test “Name” has failed. Please perform the following: 1. Check the weighing parameter setting

Page 7

Settings for Adjustment and Tests153.2.3 The MethodA Method describes the type of test to be performed and forms the core of a Test Sequence. The t

Page 8 - 3.1 Weights

Settings for Adjustment and Tests16 Warning: It is important to distinguish between the test tolerances used for the individual weight measurements d

Page 9

Settings for Adjustment and Tests173.2.3.2 EC Method for Eccentricity TestThe aim of the EC Method (Eccentricity Test) is to ensure that any deviat

Page 10 - 3.2 Test Sequences

Settings for Adjustment and Tests183.2.3.3 RP1 Method for Repeatability TestThe RP1 Method calculates the mean value and standard deviation (symbol s

Page 11

Settings for Adjustment and Tests193.2.3.4 RPT1 Method for Repeatability Test with TareThe RPT1 Method calculates the mean value and standard devia

Page 13

Settings for Adjustment and Tests20“Tolerances s”: The RPT1 Method uses two result tolerances (method tolerances) s T1 and s T2 which are applied to

Page 14

Settings for Adjustment and Tests213.2.3.6 SE2 Method for Sensitivity Test with 2 TestpointsIn contrast to SE1 Method, SE2 Method tests the sensiti

Page 15 - 3.2.3 The Method

Settings for Adjustment and Tests223.2.3.7 SERVICE Method for Reminder / Early NotificationThe SERVICE Method is a specialized Method which does not r

Page 16

Settings for Adjustment and Tests23“Early Warning Alert”: Define how early the warning should appear. E.g. set a reminder for 7 days before the batt

Page 17

Settings for Adjustment and Tests243.2.3.8 SET1 Method for Sensitivity Test with Tare and 1 TestpointThe SET1 Method tests the sensitivity of the bal

Page 18

Settings for Adjustment and Tests253.2.3.9 SET2 Method for Sensitivity Test with Tare and 2 TestpointsIn contrast to SET1 Method, SET2 Method tests

Page 19

Settings for Adjustment and Tests263.3 TasksTest Sequences define which test is to be performed and with which test weight(s). Tasks define when the Te

Page 20

Settings for Adjustment and Tests273.3.1 Task StateThe task can be switched on or off in the task status window as illustrated below. Tasks that ar

Page 21

Settings for Adjustment and Tests283.3.3 Starting MethodThe Test Sequence defined in the Task can be started in 4 different ways. Choose from:“Manual”

Page 22

Settings for Adjustment and Tests29 Note: If you change the date/time of the terminal, it is recommended to reset the previously calculated next sc

Page 23

Contents3Contents1 Introduction to Adjustments and Tests ...

Page 24

Settings for Adjustment and Tests303.3.5 Defined UsersDefine which User Profiles have the right to execute the Task. Regardless of the Starting Method u

Page 25

Settings for Adjustment and Tests313.4 ProFACT / int. Adj.ProFACT is Professional Fully Automatic Calibration Technology and provides fully automat

Page 26 - 3.3 Tasks

Settings for Adjustment and Tests32“Advanced Options”: With “Advanced Options”, you are able to further expand the sequence of the ProFACT and the in

Page 27 - 3.3.1 Task State

Settings for Adjustment and Tests33“Code to Unblock”: Here, you can define the code to be used for unblocking a blocked balance by an “As Found-”, a

Page 28 - 3.3.3 Starting Method

Settings for Adjustment and Tests343.6 Automatic Adjustment Testing Using an External Test WeightIf you want to test the adjustments using an externa

Page 29 - 3.3.4 Defined Days

Settings for Adjustment and Tests353.7 Test HistoryThe balance always records all test operations that have been made and saves them in a memory wh

Page 30 - 3.3.5 Defined Users

Settings for Adjustment and Tests36“GWP History”: When the “Show” button is pressed, a window appears with a list of the results from the Test Sequen

Page 31 - 3.4 ProFACT / int. Adj

Settings for Adjustment and Tests37Select the box to have the relevant information included i.e. when a box is ticked, the information will be print

Page 32 - 3.4.1 Advanced Options

Glossary 384 Glossary - GWP FunctionsThe following are technical terms and its definitions used in this manual.Actual Value Specified value on the W

Page 33

Glossary39Sensitivity The drift of the weight value divided by the causing load drift.Sensitivity Test Checking of sensitivity.Task Tasks define

Page 34

Introduction to Adjustments and Tests 41 Introduction to Adjustments and TestsThese Operating Instructions relate to configuring your XP Balance read

Page 35 - 3.7 Test History

Index405 IndexAAction after Failure 14Action if Failure 12Actual Value 9Actual Weight 37Adj. History 35Adj. History Selection 35Adj

Page 36 - STD C OK

Index41TTare Weight 19, 24, 25Task 26Task 01 - 12 22Task List 26Task State 27Temp. Criterion 31Temperature 37Test History

Page 39

Mettler-Toledo AG Laboratory & Weighing TechnologiesCH-8606 Greifensee, SwitzerlandTel. +41 (0)44 944 22 11Fax +41 (0)44 944 30 60www.mt.com Subje

Page 40 - 5 Index

Introduction to Adjustments and Tests51.4 The Configuration ProcessTo prepare your balance for performing routine tests and adjustments is a strai

Page 41

Accessing the Adjustments and Tests62 Accessing the Adjustments and TestsAccess the system settings by either selecting the application menu using th

Page 42

Accessing the Adjustments and Tests72.1 Overview: Settings for adjustments and testsAdjust/Test SetupTest/Adj. Weights DefineTest Sequences DefineT

Page 43

Settings for Adjustment and Tests83 Settings for Adjustments and TestsThis section describes all the menu options available for defining the parameter

Page 44 - *11781338*

Settings for Adjustment and Tests9“Name”: Weight Name can be freely defined and is intended to be an easily recognizable form for the user as an alte

Comments to this Manuals