Page view 26
Interface descriptionBBA422 / BBA425 / BBK422
27User manual 22011381A 03/05
4 Interface description
4.1 SICS interface commands
The compact scales BBA422 / BBA425 / BBK422 support the command set MT-SICS
(METTLER TOLEDO Standard Interface Command Set). With SICS commands, it is
possible to configure, query and operate the scales from a PC. SICS commands are
divided up into various levels.
For further information about the MT-SICS commands set, see MT-SICS Manual (Order
No. 22 011 459) or contact METTLER TOLEDO Customer Service.
4.1.1 Available SICS commands
Command Meaning
LEVEL 0 @ Reset the scale
I0 Inquiry of all available SICS commands
I1 Inquiry of SICS level and SICS versions
I2 Inquiry of scale data
I3 Inquiry of scale software version
I4 Inquiry of serial number
S Send stable weight value
SI Send weight value immediately
SIR Send weight value repeatedly
Z Zero the scale
ZI Zero immediately
LEVEL 1 D Write text into display
DW Weight display
K Keyboard check
SR Send and repeat stable weight value
TTare
TA Tare value
TAC Clear tare
TI Tare immediately
LEVEL 2 C2 Calibrate with the external calibration weight
C3 Calibrate with the internal calibration weight
I10 Inquire or set scale ID
I11 Inquiry of scale type
P100 Print out on the printer