20
6 Appendix
6.1 Printout examples with LC-P45 and GA42 Printers
• Functions triggered by balance • Functions triggered by printer
1)
1)
possible only with LC-P45
Function: Calibration
-- BALANCE CALIBRATION -
Date: .................
Time: .................
METTLER TOLEDO
Balance
Type: CB203
SNR: 1116150017
Weight ID: .............
Weight: 500.000 ct
Ext. calibration done
Signature:
........................
--------- END ----------
Function: Verification
With automatic insertion of date
and time, with fixed weight
(calibration weight)
----- BALANCE TEST -----
08.04.97 15:14:22
METTLER TOLEDO
Balance
Type: CB203
SNR: 1126203803
Weight ID: .............
Target: .............
Actual: 500,001 g
Diff: .............
External test done
Signature:
........................
--------- END ----------
Function: Statistics
08.04.97 15:18:55
ID 45.698-3
SNR: 1126203803
1 100.00 g
2 100.01 g
3 100.01 g
4 100.00 g
5 100.00 g
n 5
x 100.004 g
s 0.005 g
srel 0.01 %
min. 100.00 g
max. 100.01 g
dif. 0.01 g
--------- END ----------
Function: Multiplier
With rounding to the nearest 5
08.04.97 15:21:50
ID 45.698-3
SNR: 1126203803
Factor 1.650
99.99 g
* 164.9835
Function: List
Printout of the current balance
settings
--------- LIST ---------
Date: ..................
Time: ..................
METTLER TOLEDO
Balance
Type: CB203
SNR: 1126203803
SW-Ver.: 1.70 2.0
Vibr.: 2
Unit 1: g
Unit 2: g
Unit 3: g
A.Off: -
Output: Printer
Baud: 2400
Bit: 7
Parity: even
Handshake:off
--------- END ----------
Function: Calibration
With automatic insertion of date
and time
-- BALANCE CALIBRATION -
08.04.97 15:13:37
METTLER TOLEDO
Balance
Type: CB203
SNR: 1126203803
Weight ID: .............
Weight: 500,000 ct
Ext. calibration done
Signature:
........................
--------- END ----------