Configuration
The various displays have the following meanings
See also section «Operation».
Set weighing process adapter 31b to
«Animal weighing»
Use vibration adapter 31a to select the desired
time cycle (integration time + reading time):
3 sec.
_
5 sec.
7 sec.
Automatic
Set data transfer mode «S» to Auto
(see also «Configuration, Sector I-Face–»). Placing a live
animal on the weighing pan automatically initiates a weighing
cycle.
To initiate a new weighing sequence, the load on the balance
between two animal weighings must be reduced to less than 2 g for
0.1 mg balances, 10 g for 0.1 mg/10 mg and 100 mg balances,
50 g for g-balances.
Note: For DeltaRange balances select smallest unit as a refe-
rence.
Balance ready tor animal weighing
Weighing cycle in process
(integration time)
Read result (Available reading time:
Display remains stable for 35 sec)
Note: If a printer is connected, the stable weight value is printed
out automatically.
Manual
With the external transfer keys (see booklet «Technical data and
adcessories»)
With the control bar of your balance (set print command to
Prt on, see «print/transfer command» in this section):
Press control bar until Start appears.
When the control bar is released, the
weighing cycle starts.
Applications
-
,
,
_
,
_
,
_
*
,
_
,
_
Animal weighing