Section 14: Technical data and accessories
107
Technical data (limit values)
Model AX26 AX26DR AX105DR
Comparator
Maximum load 22 g 21 g 110 g
Maximum load, fine range – 3 g 31 g
Readability 0.001 mg 0.01 mg 0.1 mg
Readability, fine range – 0.002 mg 0.01 mg
Repeatability (sd) at full load
1)
0.002 mg 0.006 mg 0.05 mg
Repeatability (sd) (at gross load)
1)
0.0015 mg (1 g) 0.002 mg (1 g) 0.015 mg (10 g)
Linearity 0.006 mg 0.008 mg 0.15 mg
Eccentric load deviation at 1/2 maximum capacity
2)
0.025 mg 0.0025 mg 0.25 mg
Sensitivity offset
3)
2.5x10
-6
3.0x10
-6
4.0x10
-6
Sensitivity temperatur drift
4)
0.0001%/°C 0.0001%/°C 0.0001%/°C
Sensitivity stability
5)
0.0001%/a 0.0001%/a 0.0001%/a
Weighing time typical
6)
16 s 16 s 12 s
Interface update rate 7 /s 7 /s 7 /s
Balance dimensions (WxDxH) (mm) 241x505x293 241x505x293 241x505x293
Control unit dimensions (WxDxH) (mm) — — —
Usable height of draft shield 240 mm 240 mm 240 mm
Weighing pan dimensions (WxD) (mm) ø 28 ø 32 80 x 80
Typical data for determination of the measurement uncertainty
Repeatability (sd) typical 0.001 mg + 2.5x10
-8
• R
gr
0.004 mg + 5x10
-8
• R
gr
0.04 mg + 1x10
-7
• R
gr
Repeatability (sd) in fine range typical
7)
— 0.0015 mg + 5x10
-7
• R
gr
0.01 mg + 6x10
-7
• R
gr
Differential nonlinearity (sd) typical 5x10
-14
• R
nt
8x10
-14
• R
nt
6x10
-12
• R
nt
Diff. eccentric load deviation (sd) typical 2.5x10
-7
• R
nt
2.5x10
-7
• R
nt
5x10
-7
• R
nt
Sensitivity offset (sd) typical
3)
8x10
-7
• R
nt
8x10
-7
• R
nt
8x10
-7
• R
nt
Minimum weight
(
according to USP
) typical
8)
3 mg + 7.5x10
-5
• R
gr
12 mg + 1.5x10
-4
• R
gr
120 mg + 3x10
-4
• R
gr
Minimum weight (according to USP) in fine range typical
7)
— 4.5 mg + 1.5x10
-3
• R
gr
30 mg + 1.8x10
-3
• R
gr
Minimum weight
(@ U=1&, 2 sd) typical 0.2 mg + 5x10
-6
• R
gr
0.8 mg + 1x10
-5
• R
gr
8 mg + 2x10
-5
• R
gr
Minimum weight
(@ U=1&, 2 sd)
in fine range typical
7)
—
0.3 mg + 1x10
-4
• R
gr
2 mg + 1.2x10
-4
• R
gr
1)
Valid for compact ojects
2)
According to OIML R76
3)
After adjustment with built-in reference weight
4)
In the temperature range 10...30 °C
5)
Sensitivity drift/year after putting into operation for the first time, with the FACT self-calibration function activated
6)
Includes sample handling and setting time
7)
With DeltaRange models: Fine range starting at zero gross load
sd = Standard deviation
Rgr = Gross weight
Rnt = Net weight (sample weight)
a = Year (annum)
8)
The minimum weight can be improved by the following measures:
- Selecting suitable weighing parameters
- Choosing a better location
- Using smaller taring containers
- Using an inner draft shield (AX)
The information contained in this document has been carefully prepared and represents the most up-to-date status. Data shown as typical are reference values
for calculation of the expected measurement uncertainty. The actual measurement performance may be affected negatively or positively by the place of use and/
or the settings.