Section 5: System settings
30
5.4 Specifiying the weighing parameters
“Weighing mode”
You can use this setting to match the balance to the type of weighing. Select “Universal”
weighing mode for all normal weighing processes, or “Dosing” for dispensing liquid or
powdery weighing samples. With this setting the balance responds very quickly to
minutest changes in weight. The “Sensor mode” setting delivers a raw, unfiltered
weighing signal and is only suitable for special applications. If the “Checkweighing”
setting is used, the balance only reacts to large changes in weight, and the weighing
result is very stable.
Factory setting: “Universal”
“Environment”
You can use this setting to adapt the balance optimally to the environmental conditions
at the workplace. If you work in surroundings which are practically free from temperature
fluctuations, drafts, and vibrations, select “Very stable”. On the other hand, if you work
in surroundings where the conditions are constantly changing, select “Very unstable”.
In between these two settings there is a choice of three further settings.
Factory setting: “Standard”
“Measurement release”
You can use this setting to specify how rapidly the balance considers the measurement
value to be stable and releases it. The “Very fast” setting is recommended if you require
rapid results and their repeatability is of secondary importance. The “Very reliable”
setting gives very good repeatibility of the measurement results, but lengthens the
stabilization time. In between these two settings there is a choice of three further settings.
Note: If you choose a setting other than “Reliable + fast”, a window appears with
additional information about the selected setting.
Factory setting: “Reliable + fast”