Chapter 5 : Service and Repair
5-9 (08/04)
a. Pull processor out of the socket by the use of a chip puller. Insert chip
puller under the processor at one of the corners and pull up.
b. Position new processor above the PCB socket so the beveled corner of the
processor is lined up with the bevel corner of the socket. Carefully press
processor into socket.
10. Replace loadcell harness in reverse fashion to step 5.
11. Install display cable and communication cable (USB or RS232) to the scale
connector ports.
12. Install PCB cover, and power up scale. An Error 3 code will appear indicating the
scale needs to be reprogrammed and re-calibrated. Press UNITS key and hold to
access Setup mode. Enter required set up data and calibrate the scale per Chap 4.
Load Cell
Replacement
For load cell replacement, please use correct load cell part number equivalent to the
capacity on the data label. Contact Mettler Toledo Technical Service if you are not sure
of the correct part number.
To change out the load cell:
1. Remove power from the scale and remove platter
2. Make sure you are statically grounded with appropriate wrist wire or other means.
3. Remove spider from the load cell by:
a. PS15: Use 5 mm Hex wrench to remove two socket head bolts on top of
spider. Remove white plastic cover on spider.
b. PS60,6L,30, and 3L: Use 6 mm Hex wrench to remove two socket head
bolts on top of spider
c. PS90: To expose loadcell bolts, remove 4 small black rubber covers on
the center ridge of the top plate. Bolts are approximately 1 inch (2.5 cm)
apart from each other. Use 5 mm Hex wrench to remove the 4 socket
head bolts.
4. Remove PCB cover. Note, the scale connectors are attached to the PCB.
a. PS15 : Turn scale over, remove bolt, push on sides of plastic cover to
release the cover tabs from the base slots.
b. PS60, 6L, 30 & 3L: On the top of the scale, unscrew the bolt that holds
the PCB cover to the scale. Lift off cover.
c. PS90: Unloosen the two screws holding the connector/PCB cover to the
scale base. PCB board is attached to the cover.
5. Remove load cell harness from connector on PCB board (upper right corner – see
photo in previous section).
a. PS60, 6l, 30, 3L, and 15: Carefully pull up black tabs on each side of
connector and pull ribbon harness out.
b. PS90: Pull up on white tabs of harness connector.
6. Turn scale over to access the bottom load cell bolts, and remove load cell from the
scale base.
a. PS15: Use 5 mm Hex wrench to remove two bolts on bottom of base by
the load cell harness.
b. PS60,6L, 30, and 3L: Use 6 mm Hex wrench to remove two bolts on
bottom of scale base.
c. PS90: To expose load cell bolts, remove 4 small black rubber covers on
the center ridge under the bottom plate. Bolts are approximately 1 inch
Page view 57