3.1.2TerminalKey assignments and terminal connection.78109121312116123456Front viewDesignation Explanation1 SmartSens Two hands-free sensors. Both
9 Tare This key is used to tare the balance manually (only necessaryfor normal weighings). When the balance has been tared, theNet symbol is displayed
678541 2 39Designation ExplanationApplication name Select application.The application menu can be selected by tapping this zone. This menu can alsobe
Screen saverIf the balance is not used for 15 minutes, the display is automatically dimmed and the pixels are inverted aboutevery 15 seconds. When the
●User-specific settings●Applications●Application-specific settingsNoteA displayed menu can be left at any time by repressing the same menu key.3.2.3.1
2 Contents area The contents area is the main work area for menus and applications. The contents depend on the specific application or initiated acti
54123Designation Explanation1 Title bar The title bar displays elements for user orientation and information.2 Contents area The contents area is the
4Installation and Putting into Operation4.1UnpackingOpen the balance packaging. Check the balance for transport damage. Immediately inform a MET
1 Place the terminal (6) at the front of the balance.2 Hold the balance (7) by the guide or handle. Hold the terminal firmly with the other hand. Pul
4.3LocationAn optimal location will ensure accurate and reliable operation of the balance. The surface must be able tosafely take the weight of the
1 Insert the top draft shield door (1) at an angle (slightly below30 degrees) into the rear guide.2 Carefully fold the draft shield door (2) downwards
1 Insert the terminal support.2 Place the cable in the guide of the terminal support.3 Insert the terminal support into the opening in the front draft
Balance and terminal are at the final location.1 Connect the AC adapter (1) to the connection socket (2) atthe rear of the balance.2 Connect the AC ad
1 To start the leveling assistant, tap [LevelGuide] in the warning message.Window with level indicator is displayed in real-time.2 Observe the level
The glass draft shield doors can be opened and closed by pressing [ ], with the "SmartSens" sensors or manually .Try different combinations
5Maintenance5.1CleaningPeriodically clean the weighing pan, the drip tray, the housing, and the terminal of your balance using thebrush supplied
6Technical Data6.1General dataCAUTIONOnly use an approved AC adapter with a current-limited SELV output.Ensure correct polarity Power supplyAC
10 kΩ coupling resistor forelectrostatic dischargeInput 100…240 VAC Output 12 VDCDouble InsulationPlastic HousingPNEACDCEquivalent circuit diagram6.3
XPE205 XPE205DR XPE204Linearity deviation 0.1 mg 0.15 mg 0.2 mgEccentricity deviation (test load) 1)0.2 mg (100 g) 0.25 mg (100 g) 0.25 mg (100 g)Sens
Table of ContentsIntroduction1 5Safety Information2 6Explanation of warnings and symbols2.1 6Product safety information2.2 6Design and Functio
*30089337*For more informationwww.mt.com/xpe-analyticalMettler-Toledo AG, Laboratory WeighingCH-8606 Greifensee, SwitzerlandTel. +41 (0)44 944 22 11Fa
1IntroductionThank you for choosing a METTLER TOLEDO balance. This balance offers numerous weighing and adjustment options with exceptional operati
2Safety Information2.1Explanation of warnings and symbolsSafety notes are indicated by signal words and warning symbols and contain warnings and
Safety notesWARNINGRisk of electric shockUse only the original universal AC adapter delivered with your balance, and check that thevoltage printed on
3Design and Function3.1Overview3.1.1Balancewww.mt.comXPE204FFMETTLER TOLEDOStaticDetect14235698310 611121314715Front view1 Terminal 2 Display
67854321Rear1 Aux 1 (connection for "ErgoSens", hand or footswitch)2 Aux 2 (connection for "ErgoSens", hand or footswitch)3 RS232C