Mettler Toledo XPE Series Service Manual Page 27

  • Download
  • Add to my manuals
  • Print
  • Page
    / 32
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 26
10 k coupling resistor for
electrostatic discharge
Input 100…240 V
AC Output 12 VDC
Double Insulation
Plastic Housing
P
N
E
AC
DC
Equivalent circuit diagram
6.3Model-specific data
More detailed information is in the Operating Instructions on the CD-ROM.
XPE206DR XPE105 XPE105DR
Limit values
Maximum capacity 220 g 120 g 120 g
Readability 0.01 mg 0.01 mg 0.1 mg
Tare range (from…to) 0 … 220 g 0 … 120 g 0 … 120 g
Maximum capacity in fine range 81 g 41g
Readability in fine range 0.005mg 0.01mg
Repeatability (at nominal load) sd 0.03 mg (200 g) 0.03 mg (100 g) 0.06 mg (100g)
Repeatability (at low load) sd 0.015 mg (10 g) 0.015 mg (5 g) 0.05 mg (5 g)
Repeatability in fine range (at low load) sd 0.01 mg (10 g) 0.015 mg (5g)
Linearity deviation 0.1 mg 0.1 mg 0.15 mg
Eccentricity deviation (test load)
1)
0.2 mg (100 g) 0.12 mg (50 g) 0.2 mg (50 g)
Sensitivity offset (test weight) 0.5 mg (200 g) 0.3 mg (100 g) 0.4 mg (100 g)
Sensitivity temperature drift
2)
0.0001%/°C 0.0001%/°C 0.0001%/°C
Sensitivity stability
3)
0.0001%/a 0.0001%/a 0.0001%/a
Dimensions
Balance dimensions (W×D×H) 263 × 493 × 322mm 263 × 493 × 322mm 263 × 493 × 322mm
Weighing pan dimensions 78 × 73 mm (W × D) 78 × 73 mm (W × D) 78 × 73 mm (W × D)
Weights for routine testing
OIML CarePac
Weights
200 g F2, 10 g F1
#11123001
100 g F2, 5 g E2
#11123002
100 g F2, 5 g E2
#11123002
ASTM CarePac
Weights
200 g 1, 10 g 1
#11123101
100 g 1, 5 g 1
#11123102
100 g 1, 5 g 1
#11123102
sd = Standard deviation
1)
Valid for compact objects
2)
After adjustment with built-in reference weight
3)
On DeltaRange models: fine range starts at zero load (gross)
More detailed information is in the Operating Instructions on the CD-ROM.
XPE205 XPE205DR XPE204
Limit values
Maximum capacity 220 g 220 g 220 g
Readability 0.01 mg 0.1 mg 0.1 mg
Tare range (from…to) 0 … 220 g 0 … 220 g 0 … 220 g
Maximum capacity in fine range 81 g
Readability in fine range 0.01 mg
Repeatability (at nominal load) sd 0.03 mg (200 g) 0.06 mg (200 g) 0.07 mg (200 g)
Repeatability (at low load) sd 0.015 mg (10 g) 0.05 mg (10 g) 0.05 mg (10 g)
Repeatability in fine range (at low load) sd 0.015 mg (10 g)
27Technical Data
Page view 26
1 2 ... 22 23 24 25 26 27 28 29 30 31 32

Comments to this Manuals

No comments