PS15 Shipping Scales Technical/Operator Manual A16577800A 09/03
(09/03) 1-2 Optional Accessories • Car lighter jack • Ball transfer top platter • Tower display with 14 ft cable • Wall mount display with 14 f
Chapter 1: Introduction Electrical Interfaces (09/03) 1-3 Power Requirements The PS15 operates over an input voltage range of 7.5 to 12 VDC (at 60 mA
(09/03) 1-4 RFI Susceptibility The PS60 meets the requirements of the European Norm. 45501 for RFI susceptibility as listed below with a maximum of o
Chapter 2: Installation and Calibration Unpacking and Setup (09/03) 2-1 . 2 Installation and Calibration This chapter gives detailed instructions a
METTLER TOLEDO PS15 Shipping Scale Technical/Operators Manual (09/03) 2-2 4. Level the scale by turning the adjustable feet on the bottom of the unit
Chapter 2: Installation and Calibration Full Calibration (09/03) 2-3 WARNING! AC power sources must have proper short circuit and over current prot
METTLER TOLEDO PS15 Shipping Scale Technical/Operators Manual (09/03) 2-4 Basic Information The following sections describe some basic information
Chapter 2: Installation and Calibration Full Calibration (09/03) 2-5 The functions for each key in normal operating mode are as follows: • UNITS—Pr
METTLER TOLEDO PS15 Shipping Scale Technical/Operators Manual (09/03) 2-6 GEOCAL™ Location Codes State Code State Code State Code Alabama Kans
Chapter 2: Installation and Calibration Full Calibration (09/03) 2-7
©Mettler-Toledo, Inc. 2003 No part of this manual may be reproduced or transmitted in any form or by any means,
METTLER TOLEDO PS15 Shipping Scale Technical/Operators Manual (09/03) 2-8 The sequence for selecting your GEOCAL™ code is as follows: 1. Press the Z
Chapter 2: Installation and Calibration Full Calibration (09/03) 2-9 4. Connect the power to the scale. 5. Enter the setup mode by pressing and
METTLER TOLEDO PS15 Shipping Scale Technical/Operators Manual (09/03) 2-10 10. Press UNITS until the Build prompt appears, then press ZERO to display
Chapter 2: Installation and Calibration Full Calibration (09/03) 2-11 Metrological Seal Installation If a wire seal is required for W & M requ
METTLER TOLEDO PS15 Shipping Scale Technical/Operators Manual (09/03) 2-12 Windows 98 The set-up for the USB drivers for Windows 98 is fairly autom
Chapter 3: Configuring the Setup Parameters Configuring Setup Parameters (09/03) 3-1 3 Configuring the Setup Parameters This chapter discusses b
METTLER TOLEDO PS15 Shipping Scale Technical/Operators Manual (09/03) 3-2 There are two alternate ways to exit the setup mode: • The scale automatic
Chapter 3: Configuring the Setup Parameters Configuring Setup Parameters (09/03) 3-3 2. Press ZERO to display the desired setting, 2 pct, 5 pct, o
METTLER TOLEDO PS15 Shipping Scale Technical/Operators Manual (09/03) 3-4 Multi-interval PS15 Power-up units = Pounds Power-up units = Metric 30-n
Chapter 3: Configuring the Setup Parameters Configuring Setup Parameters (09/03) 3-5 Filter Program Block The Filter program block lets you config
METTLER TOLEDO Publication Problem Report If you find a problem with our documentation, please complete and fax this form to (864) 472-7575 Publ
METTLER TOLEDO PS15 Shipping Scale Technical/Operators Manual (09/03) 3-6 Parity Program Block The Parity program block lets you select the parity o
Chapter 3: Configuring the Setup Parameters Configuring Setup Parameters (09/03) 3-7 Sleep Program Block The Sleep program block lets you configu
METTLER TOLEDO PS15 Shipping Scale Technical/Operators Manual (09/03) 3-8 2. Press ZERO to display the desired exit option. Exit options include: •
Chapter 4: Operating Instructions Keypad and Display (09/03) 4-1 4 Operating Instructions This chapter gives information that an operator will
METTLER TOLEDO PS15 Shipping Scale Technical/Operators Manual (09/03) 4-2 1. Press ZERO to capture zero. The display reads 0.00 and a cursor appears a
Chapter 5: Service and Maintenance Cleaning and Regular Maintenance (09/03) 5-1 5 Service and Maintenance This chapter gives information on servic
METTLER TOLEDO PS15 Shipping Scale Technical/Operators Manual 5-2 (09/03) Error Code Section Error codes are displayed on the weight display with
Chapter 5: Service and Maintenance Installing the Base Mount Display (09/03) 5-3 Main PCB From PC - DB9-F RS232 SERIAL PORT To PS – RJ-12 2 RECE
METTLER TOLEDO PS15 Shipping Scale Technical/Operators Manual 5-4 (09/03) Load Cell Replacement For load cell replacement, please use current re
Chapter 5: Service and Maintenance Installing the Base Mount Display (09/03) 5-5 Installing the Base Mount Display To install the base mount displ
METTLER TOLEDO PS15 Shipping Scale Technical/Operators Manual 5-6 (09/03)
Chapter 6: Parts and Accessories PS15 (09/03) 6-1 6 Parts and Accessories Please refer to the following diagrams and parts lists when ordering pa
METTLER TOLEDO PS15 Shipping Scale Technical/Operators Manual 6-2 (09/03) PS15 Scale Parts Figure 6-a
Chapter 6: Parts and Accessories PS15 Parts List (09/03) 6-3 PS15 Parts List NOTE: (*) Some part numbers may have a revision letter in front of t
METTLER TOLEDO PS15 Shipping Scale Technical/Operators Manual 6-4 (09/03) Weight Display Parts Parts List—Weight Display Ref # Part Number
Appendix A: Host Interface Communication Parameters (09/03) Appendix-1 Appendix A: Host Interface The METTLER TOLEDO shipping scale can functi
METTLER TOLEDO PS15 Shipping Scale Technical/Operators Manual Appendix-2 (09/03)ASCII Characters and Conversions ASCII CHAR. DEC HEX 76543210 ASC
Appendix A: Host Interface Scale Status Byte Format (09/03) Appendix- 3ASCII CHAR. DEC HEX 76543210 ASCII CHAR. DEC HEX 76543210 @ 64 40 0100000
METTLER TOLEDO PS15 Shipping Scale Technical/Operators Manual Appendix-4 (09/03)Toledo Protocol Host Commands Following is a listing of host command
Appendix A: Host Interface Scale Status Byte Format (09/03) Appendix- 5 Scale Status Byte Format When communicating in host mode using the standard
INTRODUCTION This publication is provided as a guide for individuals in the operation, use, and care of this METTLER TOLEDO product. Further inform
METTLER TOLEDO PS15 Shipping Scale Technical/Operators Manual Appendix-6 (09/03) Status Byte Bit No. Bit Description 6 1 = New status data availab
Appendix A: Host Interface Scale Status Byte Format (09/03) Appendix- 76. Add the requested calibration weight to the scale, then send the command Y
METTLER TOLEDO PS15 Shipping Scale Technical/Operators Manual Appendix-8 (09/03)The table below lists the programming options available in each prog
Appendix A: Host Interface Scale Status Byte Format (09/03) Appendix- 90x06 = ASCII <ACK> 0x07 = ASCII <BEL> 0 X10 = Stop Bit: 0x01
Appendix B: USB Host Interface References (09/03) Appendix-1 Appendix B: USB Host Interface The USB 1.1 full speed (12Mbs) interface on this
METTLER TOLEDO PS15 Shipping Scale Technical/Operators Manual Appendix-2 (09/03) Protocol This interface supports the standard Universal Serial Bu
METTLER TOLEDO 1900 Polaris Parkway Columbus, Ohio 43240-2020 P/N: A16577800A (09/03) METTLER TOLEDO®
PRECAUTIONS WARNING DISCONNECT ALL POWER TO THIS UNIT BEFORE INSTALLING, SERVICING, CLEANING, OR REMOVING THE FUSE. FAILURE TO DO SO COULD RE
CONTENTS 1 Introduction...1-1 Standard Feat
4 Operating Instructions...... 4-1 Keypad and Display...
Chapter 1: Introduction Electrical Interfaces (09/03) 1-1Z 1 Introduction Thank you for purchasing a PS15 shipping scale from METTLER TOLEDO. The