Chapter 5: Service and Maintenance
Installing the Base Mount Display
(09/03) 5-3
Main PCB
From PC - DB9-F RS232
SERIAL PORT
To PS – RJ-12
2 RECEIVE 2 TRANSMIT
3 TRANSMIT 3 RECEIVE
5 SIG
GROUND
4 SIG
GROUND
Table 1 Standard PC RS232 DB9 To PS15
To test the Main PCB: (Unit must be setup for standard MT communications
(Default)):
1. Start your communications software such as ComTool (Part Number
KN000000K64).
2. Setup your software for:
7 data bits
Even parity
1 stop bit
Baud rate that matches the computer’s baud rate
Blank or Half Display
Remove power then check the display interface harness from the scale to the
display. Apply power the unit. If the blank display continues, replace the
0270 Display.
No Keypad Interaction
To test operation of the keypad, remove power, then reapply. With power to
the unit, and the W1 calibration jumper installed, attempt to enter setup mode
by pressing and holding the UNITS key. If the display does not indicate setup
mode, replace the 0270 Display.
Display Locked
To test operation of the display, remove power, then reapply. With power to
the unit, add weight. If no motion is displayed, replace either the load cell or
the Logic PCB.