Chapter 3: Configuring the Setup Parameters
Configuring Setup Parameters
(09/03) 3-3
2. Press ZERO to display the desired setting, 2 pct, 5 pct, or 100 pct.
3. Press UNITS again to accept the displayed option. The PS15 continues to
the Zero Cursor program block.
Zero Cursor Program
Block
The Zero Cursor program block lets you enable or disable the center of zero
display on the PS. (Metrology PCB jumper must be removed.)
To configure the program block:
1. Press UNITS to display the 0 CurS prompt, then press ZERO.
2. Press ZERO to display the desired approval setting, enable or disable.
3. Press UNITS to accept the displayed option. The PS15 continues to the
Power-up Unit program block.
Power-up Unit
Program Block
The Power up Unit program block lets you select which units will be active on
the PS upon startup. This program block will effect the build options that are
seen in the Build program block.
To configure the program block:
1. Press UNITS to display the unitS prompt, then press ZERO.
2. Press ZERO to display the desired approval setting, pounds or metric.
3. Press UNITS to accept the displayed option. The PS15 continues to the
Build / Alternate Units program block.
Build Program Block
The build program block lets you chose a capacity and increment size setting
for the product based on the model you have purchased and the power-up
units selection. (Metrology PCB jumper must be removed.)
To configure the program block:
1. Press UNITS to display the buiLd prompt, then press ZERO.
2. Press ZERO to display the desired approval setting. Option may vary as
follows:
Standard PS15
Power-up units = Pounds
Power-up units = Metric
30-01 30 x 0.01lb.* 15-005 15 x 0.005kg
30-005 30 x 0.005lb. * 15-002 15 x 0.002kg