Appendix A: Host Interface
Scale Status Byte Format
(09/03) Appendix-
7
6.
Add the requested calibration weight to the scale, then send the command
Y to continue. The scale responds <STX> CAL DONE <CR><LF> or
<STX> INCORRECT AMOUNT OF WEIGHT <CR><LF>.
Configure Scale
Parameters Using
Host Interface
Configuration parameters can also be sent using the host interface. A
command to send a new scale configuration over the host interface is available
in the METTLER TOLEDO command set.
To configure using the host interface:
1. From the host computer, send the host command S to initiate the
configuration setup sequence. The scale responds with the current set
bytes:
<STX>X
1
X
2
X
3
X
4
X
5
X
6
X
7
X
8
X
9
X
10
X
11
X
12
X
13
X
14
X
15
X
16
X
17
X
18
X
19
X
20
<CR>
2. Determine the command string representing the configuration parameters
that are desired. The command string consists of 22 characters beginning
with an ASCII <STX> and terminated with an ASCII <CR>. The
command string represents a 10-byte bit-mapped setup array stored in
EEPROM. The 10-byte array is expanded for serial transmission so that
the high-order four bits of each byte are zeroes and the low-order four bits
are numbers representing the desired option for each program block. See
the table below for definitions of each byte.