Mettler Toledo GA42 Operations Instructions Page 17

  • Download
  • Add to my manuals
  • Print
  • Page
    / 44
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 16
METTLER TOLEDO DL36 15
Short-form report GLP report On printout:
Date 96/06/02 13:26 Date 96/06/02 13:26 Date and time
Sample ID ETHANOL Instr. ID DL36-82 Instrument identification
Weight 0.1243 g Sample ID ETHANOL Sample identification
Result 219.9 µg Weight 0.1243 g Weight
0.17689 % Factor 1.00 Factor
Result 219.9 µg Result
0.17689 %
Titr.Time 00:00:26 Titration time
Drift 1.62 µg/min Drift value
Blank 0 µg Blank value
User ID KVE User identification
Key selection Display
Result
Calculation
Select
Report
Enter
Short
Select
GLP
Enter
Report
Description
Select either the short-form or GLP format.
To titrate, press Esc.
Calculation formulae
Formula No. Result in Formula
1 µg or mg
2 ppm or %
(entry of the
sample weight)
T
value
: Result of the titration in µg t: Titration time in min (see page 13)
drift: Drift value in µg/min blank: Blank value in µg
weight: Sample weight in g
2. Report parameters
You can choose between two report formats: short-form and GLP format (GLP: Good
Laboratory Practice).
The method
factor T d ift
value
r t blank)⋅⋅(
factor
T d ift
weight
value
r t blank)
(
Page view 16