6 METTLER TOLEDO DL36
Putting into operation
4. Putting into operation
4.1 Preparing the titration cell
Never inhale the vapors of Karl Fischer reagents and avoid skin contact!
The reagents could hazard your health.
Remove the Teflon stopper of the titration cell and carefully slip the magnetic stirrer into
the cell; replace the stopper.
Place the titration cell in the holder.
When the DL36 is delivered, the glass joints for the double-pin platinum electrode, inner
burette and Teflon stopper have a very thin coating of silicon grease to assure tightness.
Remove the Teflon stopper and add anolyte up to the lower mark on the titration cell
(100 mL) using the funnel.
Remove the silicon stopper from the inner burette and add 5 mL catholyte to the cathode of
the inner burette; its level should be just below that of the anolyte.
Check that the glass joint is greased, then insert the drying tube.
Plug the cables of the electrode and the inner burette into the appropriate inputs at the rear
of the DL36 (see illustration on next page).
WARNING
Glass stopper with inner burette
and silicon stopper/drying tube
Drying tube
Titration cell
Detection electrode (double-
pin platinum electrode)
Keypad
Holder of titration cell
Display
Teflon stopper with septum
(injection port)