26 METTLER TOLEDO DL36
Error messages and malfunctions
10. Error messages and malfunctions
10.1 Error messages of the DL36
Error message
A.Capacity Over!
C.Capacity Over!
Current error!
Elect open!
Elect short!
Meas. Over!
Over Titr.!
Para set miss!
Pre Amp Err-XX
Cause
The summed electrolytic current
(converted to mg H
2
O) of the anolyte
is greater than its capacity
The summed electrolytic current
(converted to mg H
2
O) of the catholyte
is greater than its capacity
There is no current flowing between
anode and cathode
The detection electrode is not working
properly (break)
The detection electrode is not working
properly (short circuit)
Measurement range exceeded. The
amount of moisture in this sample is
greater than 100 mg H
2
O
Overtitrated: the anolyte contains an
excess of iodine
Neither of the two abort parameters
t(max) and drift stop is defined
Error in the preamplifier
10.2 Other malfunctions
Fault
Instrument can not
be switched on
Malfunctions of
the stirrer
Possible causes
No line voltage
Power cable not connected
Fuses faulty
Stirrer speed set to 0
Measures
Check power cable, on/off switch, voltage
selector and fuses
Check stirring speed (stirrer key and size
of the magnetic stirrer (35 mm)
Measures
• Change anolyte
• Set capacity of the anolyte to 0 (see
Section 8.1)
• Change catholyte
• Set capacity of the catholyte to 0 (see
Section 8.1)
• Check anolyte and catholyte
• Check supply leads of cathode and
anode
Plug in electrode correctly
Rectify short circuit (align platinum pins
correctly)
Change catholyte, inject smaller amount
of sample
• Clean anode
• Protect instrument against direct
sunlight
Define either t(max), the drift stop or
both parameters
Contact METTLER TOLEDO Service
Comments to this Manuals