28 METTLER TOLEDO DL36
11. Maintenance and servicing
11.1 Charging internal battery
A battery is responsible for the internal clock. Contact the METTLER TOLEDO Service when it
is discharged, i.e. when the date and time are no longer correctly displayed.
11.2 Line voltage, line fuses
The DL36 can operate in different voltage ranges. The voltage selector is used to select the
range.
Voltage range/supply Setting on voltage selector Fuses
100 V, 50/60 Hz 100 VAC
T3,15L250V
120 V, 50/60 Hz 120 VAC
220 V, 50/60 Hz 220 VAC
T1,6L250V
230/240 V, 50/60 Hz 230/240 VAC
Changing voltage/fuses
– Switch the instrument off and disconnect the power cable!
Open the cover above the power cable socket using a flat screwdriver.
Take out the drum of the voltage selector and reinsert horizontally with the
desired voltage.
Remove both fuse holders using a screwdriver and insert the appropriate
fuses (see above).
Reinsert the fuse holders (note direction of arrow) and replace the cover.
The same procedure applies to the replacement of blown fuses.
Maintenance and servicing
Voltage selector
Fuse holder
Flat srewdriver
Voltage selector
(indicator)