Attention:
Do not place the balance on the pan support location bolt (0.1 mg and 1 mg models).
Models with a glass draft shield: Carefully lift the draft shield from the weighing platform and
put it aside.
Note:
For below-the-balance weighing with the L Platform models, you will need hook 11132565
from the accessories range.
Weighing below the balance is not possible with "MS-KL" models.
S Platform L Platform
A
B
B
A
a) Switch off the balance and remove the power cable and any interface cable from the balance.
b) Remove the draft shield element (10 mg models).
c) Remove the weighing pan and pan support.
d) Remove the bottom plate and unlock the draft shield (models with draft shield).
e) Remove the 2 screws (A) and the cover plate (B). The hanger is now accessible.
f) Then turn the balance to its normal position and simply reinstall all components in the reverse order.
4.8 Adjustment (Calibration)
To obtain accurate weighing results, the balance must be adjusted to match the gravitational
acceleration at its location. Adjusting is necessary:
before the balance is used for the first time.
at regular intervals during weighing service.
after a change of location.
4.8.1 Fully Automatic Adjustment FACT
Note: On models with FACT only.
The factory setting is fully automatic adjustment FACT (Fully Automatic Calibration Technology) with the internal
weight (see also section "The Menu"). In this setting, you have no need worry about adjusting your balance.
The balance adjusts itself automatically:
after the warm-up phase on connection to the power supply.
MS-S / MS-L Models20
Setting up the Balance