
S I Command not executable (balance is currently
executing another command, e.g. taring, or time
out as stability was not reached).
S + Balance in overload range.
S - Balance in underload range.
Example
Command S Query a stable weight value.
Response S S 100.00 g The current stable weight value is 100.00 g.
The available MT-SICS commands are listed in the table. For further information please refer to the Reference Manu
al "MT-SICS 11780711" downloadable from the Internet under www.mt.com/sics-newclassic.
Description Description
@ Cancel (Reset) M31 Mode after restart
C0 Query/Set adjustment settings M46 Print interval
C1 Start adjustment according to current settings PW Piece counting: Query/Set piece weight
C2 Start adjustment with external weight PWR Power on/off (PWR 0 means switch off com
pletely, if balance is powered by battery)
C3 Start adjustment with internal weight S Send stable weight value
D Display text sent to balance SI Send weight value immediately
DAT Date query/set SIR Send weight value immediately and repeat
DW Display weigh SIRU Send weight value with currently displayed
unit immediately and repeat
I0 Commands implemented SIU Send weight value with currently displayed
unit immediately
I1 MT-SICS level and MT-SICS versions SM0 Dynamic weighing: cancel all SMx com
mands
I2 Balance data SM1 Dynamic weighing: Start immediately and
send the result
I3 Software version, type definition number SM2 Dynamic weighing: start after a minimum
load is exceeded and send result
I4 Query serial number (SNR) SM3 Dynamic weighing: start after a minimum
load is exceeded, send result and repeat
I5 Query SW-identification number SM4 Dynamic weighing: query/set time interval
I10 Query balance ID SNR Send stable weight value and repeat on
weight change
I11 Query balance type SNRU Send stable weight valuewith currently dis
played unit and repeat on weight change
I14 Query balance information SR Send weight value on weight change
K Keys: set configuration SRU Send stable weight value with currently dis
played unit on weight change
M02 Query/set environment ST Send stable weight value on pressing (print)
key
M03 Query/set AutoZero SU Send stable weight value with currently dis
played unit
M08 Display brightness T Tare
M09 Display contrast TA Get/Set tare weight value
M11 Beeper: Query/set volume TAC Clear tare value
M14 List available language TI Tare immediately
MS-S / MS-L Models72
Interface Specification
Comments to this Manuals