11 Application "Formulation" (Net Total Formulation)
The "Formulation" (Net Total) application allows you to
●
weigh in (add and store) up to 999 individual component weights and displays the total. If
a printer is connected, the component weights are printed individually and as a total.
●
tare and store up to 999 container weights and displays the total. If a printer is connected,
the tare weights are printed out individually and as a total.
Requirement: The function "FORMULA" must be assigned to an «Fx» key (see advanced menu
topic "ASSIGN:Fx"). Connect a printer or a PC if present.
CLEAR M?
No
FORMULA
M
.
CLEARED
a) Activate function formulation "FORMULA" by pressing and holding the
appropriate assigned «Fx» key.
b) Press « » to continue formulation weighing. For a new formulation
press « » to select "Yes" and press « » to clear the memory.
Note: If the memory is already cleared (sample counter is 0) the
memory clear question will be not displayed.
#
-T1-
0
.
00
g
Net
10
.
33
g
C
0
.
00
g
#
Tare container (if used):
a) Press « » to zero the balance if needed.
b) Place the empty container on the weighing pan.
c) Press « » . The container is tared and the tare count "- T1 -"
is displayed and the tare weight is printed.
Note: If you pre-tare via MT-SICS (e.g. bar code reader) "- PT1 -" is dis
played.
Weighing the first component weight:
a) Load the first component weight.
b) Press « ». The display briefly shows the component count "- 1 -" ,
the current weight is stored as sample and the component weight is
printed. The display is set back to zero.
Weighing further component weights:
The same procedure as for the first component weight with the same or
new container).
●
1...999 sample values are possible.
●
max 999 tare values are possible.
Results:
Press « », the results are displayed and printed.
MS-S / MS-L Models 47