METTLER TOLEDO DE40/50/51 Density Meter 21
5.2.3 Performing the air adjustment
Whether you have a sampling pump ASU-DE or not, the following steps will be the same:
- Press the <CALIB.> and follow the directions on the display:
If the cell is really clean and dry,
press <ENTER>.
While adjusting to the air value the
display shows:
Wait until the next message is
displayed.
5.2.4 Performing the water adjustment
- Open a vial of certified water (one of the five delivered water standards). This water will be
used for the adjustment.
Note: you can also use bi-distilled water, but we recommend to use our certified water for
traceability.
a) If you have a sampling pump ASU-DE:
- Remove the sampling tube from the measuring cell.
- Remove the drain tube connection from the ASU-DE pump (white connector with black ring).
b) The next steps will be the same wether you have a sampling pump ASU-DE or not:
- Pierce the septum of the certified water standard with the tip of the syringe.
- Fill the syringe with approx. 8 mL of certified water.
- Remove the tip of the syringe and insert the syringe in the sample inlet (lower measuring tube
connection).
- Perform an analytical rinse, i.e. fill approx. 2 mL carefully, wait for 10 s without removing the
syringe, then fill in again 2 mL very carefully. During the filling watch carefully that no air
bubbles enter the measuring cell.
Getting started
9.ADJUST 01-01
Temp.(°C) Oscillation
Set 20.00
Cell 20.00
Purge OK?
845393
9.ADJUST 01-01
Temp.(°C) Oscillation
Set 20.00
Cell 20.00
Set Water
845393
9.ADJUST 01-01
Temp.(°C) Oscillation
Set 20.00
Cell 20.00
Calibrating
845393