6 METTLER TOLEDO DE40/50/51 Density Meter
1. Introduction
The DE40 and DE50/DE51 Density Meters measure the density d of gases, liquids and
solutions using the oscillating body method. They are distinguished by their temperature
range and their measurement accuracy. All three instruments:
come with five certified water standards.
measure gases, liquids and solutions with densities (d) in the range 0 to 3 g/cm
3
.
(DE40: with 0.0001 g/cm
3
; DE50/DE51: with 0.00001 g/cm
3
).
need a minimum amount of substance (1.2 millilitres) for measurement.
keep the temperature of the sample being measured constant thanks to a built-in thermostat
control (DE40: between 4 and 90 °C, DE50/DE51: between 4 and 70 °C).
have 10 methods stored in their memory, which you can change to suit your measurements.
show the density relative to water (specific gravity) at 4 °C (d
t
4
) or at the measurement
temperature (d
t
t
).
calculate the concentration of a solution using tables or entered data of measured standard
solutions.
calculate the density of a solution for a desired temperature using an entered temperature
function.
have a built-in check function that allows to periodically check the accuracy of the instrument
without changing the instrument settings.
store a maximum of 100 measured values.
perform statistical evaluations (mean value, standard deviation, relative standard deviation)
and allow the recalculation of results.
use Memory Cards for data storage, sample files and method transfer.
Special Application Memory Cards are available for all three instruments which allow the
automatic conversion of the experimental data into BRIX degrees, or alcohol concentration.
A special API software converts the density of oil products measured at elevated
temperatures automatically into the density at 15 °C (or 60 °F).
You can use an automatic sampling unit to perform continuous measurements; the built-in
RS-232C interface allows attachment of a computer to transfer experimental data. Print out
the entered parameters and data as well as results on a printer (option).
You can connect your instrument to a RE40 or RE50 refractometer to perform simultaneous
measurements of density and refractive index with one sample.
Connection of a SC24 sample changer allows fully automated measurements, even in
combination with a refractometer (RE40 or RE50).
Note: Examples of the density given in these Operating Instructions are for the DE50/DE51.
Introduction