METTLER TOLEDO DE40/50/51 Density Meter 67
Repeat the previous steps if you
want to load more methods, being
careful not to overwrite an already
defined method: select a free
method “Method-x” to be
overwritten.
Select “Exit” with the key. Confirm
with <ENTER>.
12.8 Save sample file
With the Data Memory Card you can save sample files to the card or load sample files from
the card. The DE40, DE50 and DE51 store up to 498 sample files in the internal memory.
These sample files are usually used together with the sample changer, in order to run
samples according to a predefined sequence (sample names, method to be used, etc...).
They can also be used as a SOP (Standard Operating Procedure) if no sample changer is
connected. See section 6.1 for the description of sample files. Let’s assume that you have
already defined such a sample file.
Press the <FUNC> key.
Select “1.Sample File”. Confirm with
<ENTER>.
Select “On” with the key. Confirm
with <ENTER>.
Press the <SAMPLE> key.
Select “Save” with the key.
Confirm with <ENTER>.
< Method File >
No. Name Temp.(°C) Result .
0. Method-0 20.0 d
1. ORANGE 20.0 d
2. Method-2 20.0 d
3. Method-3 20.0 d
Load Save Exit
< Method File >
No. Name Temp.(°C) Result .
0. Method-0 20.0 d
1. ORANGE 20.0 d
2. Method-2 20.0 d
3. Method-3 20.0 d
Load Save Exit
Data Memory Card
< Function >
0.Data File 5.Set Check
1.Sample File 6.Periodic
2.Method File 7.Auto Start
3.Changer 8.CARD utility
4.Factor 9.Memory Clear
< Sample File >
Sample File : Off On
Exit : [Execute]
< Sample >
Sample File :Skip Load Save
Max.Meas.No. :[ 24]
Next Meas.No. :[ 1]
Method :Current Variable
Set Parameter :[Execute]
Exit :[Execute]