METTLER TOLEDO DE40/50/51 Density Meter 37
Concentration
Select "Conc." with the key.
Confirm with <ENTER>.
You will select this option if you want to determine the concentration of two-component
solutions, e.g. aqueous solutions of glycerol or acetone. Ask your local representative for
available concentration tables or download them from our Internet Homepage:
www.density.com.
There are two ways to program a concentration function, either by entering this function
directly (provided that this function is already known) or by entering a concentration table
(data pairs of density and corresponding concentration).
First you have to select the unit for your concentration. This unit should be the same as that
of the solution used: the instrument does not convert units.
Select your "Conc.Unit" with the
key. Confirm with <ENTER>.
Following units are available: No
unit, %, ppm,
mg/g, mg/mL, mg/L,
g/L, Eq/L, mol/L,
g/cm
3
.
Then you have to select the concentration formula. If your concentration function is known
already, select the same function. If your concentration function is not known (input of
concentration table) select the formula which best fits your concentration. As a rule of thumb
we recommend to select A+Bx+Cx
2
if you don't know which formula to select.
Case 1: Direct input of concentration function (known concentration function)
We will describe this function using an example: measurement of Apparent Extract in beer.
The function to convert density into Apparent Extract is:
Apparent Extract = -460.234 + 662.649 x d
20
20
- 202.414 x [d
20
20
]
2
The following steps have to be performed:
Select "%" as Conc.Unit with the
key. Confirm with <ENTER>.
Select "A+Bx+Cxx" with the key.
Confirm with <ENTER>.
Calculation parameters
Method 0 < Result >
Result :<Conc. >
Exit :[Execute]
Method 0 < Result >
Result : <Conc. >
Conc.Unit : <None >
Conc.Formula : <A+Bx >
Parameter Set : Table Coefficient
Data Replace : <x<-d-R >
Method 0 < Result >
Result : <Conc. >
Conc.Unit : <% >
Conc.Formula : <A+Bx+Cxx >
Parameter Set : Table Coefficient
Data Replace : <x<-d-R >