Mettler Toledo GA42 Operations Instructions Page 27

  • Download
  • Add to my manuals
  • Print
  • Page
    / 96
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 26
METTLER TOLEDO DE40/50/51 Density Meter 27
Method:
Here you can select if you want to perform all measurements with one method only (current
method) or with different methods (variable).
Example:
You always get a series of 12 samples of grape juice and 7 samples of orange juice to
measure. You want to check for each series your instrument with a check standard.
You will have to set up 3 different methods as follows:
Method-0 Method-1 Method-2
Method Name Grape.J CheckStd Orange.J
Meas.Temp. 20.00°C 20.00°C 20.00°C
Stability 2 0 2
You will the define a sample file as follows:
This sample file will run all 12 Grape
juices with method 0 "Grape.J",
then the check standard with
method 1 "CheckStd" (this method
has a higher stability as it is required
to check the instrument),
then run all 7 orange juices.
It will allow to run the statistics on
each sample separately.
If the next series would be without
the first 12 grape juices, you would
have to select:
Max.Meas.No: 20
Next.Meas.No: 13
< Sample >
Sample File :Skip Load Save
Max.Meas.No. :[ 20]
Next.Meas.No :[ 1]
Method :Current Variable
Set Parameter :[Execute]
Exit :[Execute]
Sample and sample file
< Sample List>
No. M S.No. Sample ID
1 0 01-01 GRAPE
2 0 01-02 GRAPE
3 0 01-03 GRAPE
4 0 01-04 GRAPE
5 0 01-05 GRAPE
6 0 01-06 GRAPE
7 0 01-07 GRAPE
8 0 01-08 GRAPE
9 0 01-09 GRAPE
10 0 01-10 GRAPE
11 0 01-11 GRAPE
12 0 01-12 GRAPE
13 1 50-01 STANDARD
14 2 02-01 ORANGE
15 2 02-02 ORANGE
16 2 02-03 ORANGE
17 2 02-04 ORANGE
18 2 02-05 ORANGE
19 2 02-06 ORANGE
20 2 02-07 ORANGE
Page view 26