36 METTLER TOLEDO DE40/50/51 Density Meter
9. Calculation parameters (password protected)
This section describes how to select the type of results (density, specific gravity,
concentration), the number of decimals for these results and how to set the parameters for
the concentration calculation. These calculation parameters can be set for each of the 10
methods.
Press the <CALC.PARA> key.
9.1 Density: d, d(t/4), d(t/t)
Select CALC.PARA - Density
Select what density should be
displayed and printed:
d: mass density (real density).
d(t/4): relative density referred to
water at 4 °C.
d(t/t): relative density referred to
water at the measurement
temperature.
At 20 °C: d(t/t) = d
20
20
and d(t/4) = d
4
20
9.2 Result (Density, Concentration, BRIX, ALCOHOL, USER TABLE)
Select CALC.PARA - Result.
Select which result should be
displayed and printed:
d: density, resp. d/t/t) or d(t/4),
depending on the selected
density (previous step).
Conc.: concentration.
Note: here you can also select concentration applications, if they have been previously
loaded with a Memory Card (See also section 13 "Application Memory Cards):
BRIX, ALCOHOL, USER TABLE.
Calculation parameters
Method 0 < Calc.Parameter >
0.Density
1.Result
2.Temperature Compensation
3.Decimals
Method 0 < Density >
Density :d d(t/4) d(t/t)
Exit :[Execute]
Method 0 < Result >
Result :<d >
Exit :[Execute]