26 METTLER TOLEDO DE40/50/51 Density Meter
6. Sample (Number and ID) and Sample File
Here you can define your sample identifications like sample number and sample ID.
The sample number is composed of two parts: for example [03]–[01].
[03]: The first two-digit number – ranging from 00 to 98 – represents the series number of
your measurement series. In the instrument it is called "High Sample No.". If you group your
measurements of different sample types by this series number, you can conveniently
process your data statistically later on.
[01]: The second two-digit number – ranging from 00 to 99 – is automatically increased by
one per measurement.
Additionally you can define a sample ID, for example a sample name or the Lot number.
Press the <SAMPLE> key.
Enter the sample ID using the alpha-
numeric keys (max. 10 characters
including "space", "period" and
"hyphen").
(*) Appears with DE51, or with DE40
with viscosity correction (memory
card).
6.1 Sample File (sample changer)
If you have connected a sample changer (see Function 3 "Changer") or set a sample file (see
Function 1 "Sample File") another display will appear when pressing the <SAMPLE> key:
The first line is usually skipped,
except when you want to SAVE OR
load a sample file with a Memory
Card (see section 12.8 to 12.10).
Max.Meas.No:
This is the total number of samples to be measured. This number can be higher than 24
(when used with the SC24). This number can be changed during the measurement. For
example set this number to 12 if you have only 12 samples to measure when you start and
change it later to 25 if you get 13 more samples to measure.
Next.Meas.No:
This is the next measurement number (according to the sample file). It is increased
automatically after each measurement.
< Sample >
Sample No. : [03]-[01]
Sample ID : [ORANGE ]
Viscosity : off on (*)
Exit : [Execute]
< Sample >
Sample File :Skip Load Save
Max.Meas.No. :[ 24]
Next.Meas.No :[ 1]
Method :Current Variable
Set Parameter :[Execute]
Exit :[Execute]
Sample and sample file