Page view 24
METTLER TOLEDO DE40/50/51 Density Meter 25
0.TestMeth ORANGE 02-01
Temp.(°C) Density(g/cm3)
Set 20.00
Cell 20.00
Set Rinse-1
0.99820
Getting started
5.3.6 Cleaning and rinsing of the cell with ASU-DE
Follow the instructions on the
display for the rinsing and purging.
Use a highly volatile solvent (e.g.
acetone or ethanol) for the rinsing.
5.3.7 Cleaning the cell without ASU-DE
- suck back the sample into the syringe.
- flush the measuring cell with appropriate solvent.
- if necessary flush again with a highly volatile solvent (for example acetone or ethanol) to
facilitate the drying of the cell.
- dry the measuring cell according to section 5.2.2.