METTLER TOLEDO DE40/50/51 Density Meter 45
Functions
Press the <CALC.PARA> key.
Select "1.Result" and confirm with
<ENTER>
Select Brix with the key and
confirm with <ENTER>
Select Function 0.Data File again
The recalculated result is shown
(from density to BRIX)
10.1.3 Data Deletion
This function allows to exclude single measured values from the statistics calculation by
marking them with an asterisk (they are not deleted from the list).
Select the data to be excluded with
the <UP> and <DOWN> keys.
Exclude this data by pressing the
<DISPLAY> key.
The excluded data is marked with an
asterisk.
0.Method-5 01-02
Temp.(°C) Density(g/cm3)
Set 20.00
Cell 20.00
Recalculation
1.03980
Recalc. < Calc.Parameter >
0.Density
1.Result
2.Temperature Compensation
3.Decimals
Recalc. < Result >
Result :<Brix >
Exit :[Execute]
< Data File >
No. M.Name Sample Result .
11 Method-5 01-01 d 1.03981
12 Method-5 01-02 Br 10.40
13 Method-6 02-01 Br 10.81
14 Method-6 02-02 Br 10.82
Statis. Recalc. Load Save Exit
< Data File >
No. M.Name Sample Result .
1 Method-0 01-01 d 10.81
2 Method-0 01-02 Br 8.40
3 Method-0 01-03 Br 10.81
4 Method-0 01-04 Br 10.82
Statis. Recalc. Load Save Exit
< Data File >
No. M.Name Sample Result .
1 Method-0 01-01 d 10.81
* 2 Method-0 01-02 Br 8.40
3 Method-0 01-03 Br 10.81
4 Method-0 01-04 Br 10.82
Statis. Recalc. Load Save Exit