Mettler Toledo GA42 Operations Instructions Page 55

  • Download
  • Add to my manuals
  • Print
  • Page
    / 96
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 54
METTLER TOLEDO DE40/50/51 Density Meter 55
11.1.2 Printer setup
We recommend to select the GA42 printer from METTLER TOLEDO. Make sure that the DIP
switch 2 is ON on the GA42 printer.
The following printers are predefined:
GA-: for GA42 and SPRINTER printer from METTLER TOLEDO.
Other: for all serial printers. Selection of:
- Baud rate (300, 600, 1200, 2400,....)
- Parity (none, even, odd)
- Stop bits (1, 2)
- Data bits (7, 8)
These parameters have to be set according to the specifications of
your printer.
The bold settings are active for the GA42 printer.
IDP-: not used
DP-: not used
11.2 Date and Time (Setup 1)
This option is used to enter the actual date and time.
Press the <SETUP> key.
Select "1.Date&Time". Confirm with
<ENTER>.
Actual date and time are displayed.
Press <ENTER> key.
Enter date and time (the date format
is selected in the menu Setup
„5.International”) and confirm with
<ENTER>.
Setup
< Printer >
Printer :IDP- DP- GA- Other
Exit :[Execute]
< Setup >
0.Interface 5.International
1.Date & Time 6.Lock
2.Name 7.Beep
3.Serial/Version 8.Temperature
4.LCD Contrast
< Date & Time >
12/07/1998 17:12
Day :[12] Hour :[17]
Month :[ 7] Minutes :[12]
Year :[1998] Exit :[Execute]
Page view 54
1 2 ... 50 51 52 53 54 55 56 57 58 59 60 ... 95 96

Comments to this Manuals

No comments