Message Type Meaning Remedial action
Abort I Function (taring, zero setting, weight transfer,
calibration, test, start in dynamic weighing, etc.)
interrupted by keystroke
Cal done I Calibration (adjustment) completed successfully
Illegal value F Inputted value or value calculated in piece counting Effect entry in admissible
"Count" outside admissible range, old value remains region (section 8.2.2)
unchanged
InitCal done I Adjustment of internal calibration (adjustment) finished Wait for following calibration
with internal weight
MinChng-Err F Weight can not be transferred as the load change Induce load change by
since the last weight transfer has not attained the displacement of the weighing pan
"MinChange" value (see appropriate application,
section 6.1)
no Prog I Program cassette missing or not inserted properly Insert program cassette
(properly) (section 8.7)
No reference F No piece weight or target weight available Enter piece or target weight
(see piece counting and +/–statistics, section 5)
Not allowed F On zeroing of certified balances: Zeroing is allowed Unload balance or reduce load
only up to ±2 % of the weighing range.
n = Max n I Specified number of samples (n=...) reached, To close, unload balance; to
after unloading of the balance the weighing series is determine additional samples,
automatically closed and the results cleared increase Max n or set it to zero
n > 0 –> Clear I Weighing series in progress (item counter > 0), First close weighing series
entry into the menu or alteration of the reference with "Clear"
or switching between units in +/–statistics not allowed
PassWord = I Entry into MENU protected by a password Enter valid password
Press ¥ F On taring: The balance was first switched on or First rezero balance after
zeroed with load and then unloaded. unloading
Printing I A list of the settings has been sent to the printer
(section 6.4.1)
RefOpt-Err F Piece count since the last piece weight determination Reduce piece number
more than doubled, reference optimization not allowed on the balance
Reset done I Resetting of the parameters to the factory setting
finished
8.3 Messages in the display
During operation of the balance a message or an unusual display may appear briefly in the top line of the display. There are two
different types of messages: Information (type "I") or error message if operating errors occur (type "F").