108
PR/SR balancesAppendix
Message Type Meaning Remedial action
Abort I • Function (taring, zero setting, weight transfer,
calibration, test, start in dynamic weighing, etc.)
interrupted by keystroke
Cal done I • Calibration (adjustment) completed successfully
Illegal value F • Inputted value or value calculated in piece counting • Effect entry in admissible
"Count" outside admissible range, old value remains region (section 8.2.2)
unchanged
InitCal done I • Adjustment of internal calibration (adjustment) finished • Wait for following calibration
with internal weight
MinChng-Err F • Weight can not be transferred as the load change • Induce load change by
since the last weight transfer has not attained the displacement of the weighing pan
"MinChange" value (see appropriate application,
section 6.1)
no Prog I • Program cassette missing or not inserted properly • Insert program cassette
(properly) (section 8.7)
No reference F • No piece weight or target weight available • Enter piece or target weight
(see piece counting and +/–statistics, section 5)
Not allowed F • On zeroing of certified balances: Zeroing is allowed • Unload balance or reduce load
only up to ±2 % of the weighing range.
n = Max n I • Specified number of samples (n=...) reached, • To close, unload balance; to
after unloading of the balance the weighing series is determine additional samples,
automatically closed and the results cleared increase Max n or set it to zero
n > 0 –> Clear I • Weighing series in progress (item counter > 0), • First close weighing series
entry into the menu or alteration of the reference with "Clear"
or switching between units in +/–statistics not allowed
PassWord = I • Entry into MENU protected by a password • Enter valid password
Press ¥ F • On taring: The balance was first switched on or • First rezero balance after
zeroed with load and then unloaded. unloading
Printing I • A list of the settings has been sent to the printer
(section 6.4.1)
RefOpt-Err F • Piece count since the last piece weight determination • Reduce piece number
more than doubled, reference optimization not allowed on the balance
Reset done I • Resetting of the parameters to the factory setting
finished
8.3 Messages in the display
During operation of the balance a message or an unusual display may appear briefly in the top line of the display. There are two
different types of messages: Information (type "I") or error message if operating errors occur (type "F").
Comments to this Manuals