Mettler Toledo PR/SR Operations Instructions Page 19

  • Download
  • Add to my manuals
  • Print
  • Page
    / 124
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 18
17
PR/SR balances Startup procedure
-BALANCE TEST
----BALANCE TEST---
-0.03-g-D:
1999.98 g
0.00 g
Test Calibration Repro
Start
- BALANCE TEST
AutoCal BPTGNet
17.06.95
METTLER TOLEDO
Type:
SNR:
Bal:
Weight ID:..............
Target :
Actual :
Diff :
External test done
Signature:
........................
PR5002
1114051375
Your entry
1999.98 gaa
1999.95 gaa
-0.03 gaa
----- BALANCE TEST -----
--------- END ----------
11:27:16
- BALANCE TEST
- BALANCE TEST
Cal / Test
0
Menu
Procedure for testing the balance with external weights
Requirement
The key word "External" is marked under "CAL -> Test" in the menu (see
section␣ 6.3).
Initiate test by pressing and holding the key and by pressing the SmartBar
under the word "Test".
During the test, "BALANCE TEST" appears in the bottom line of the display.
After a short time, the balance flashes the preset target weight it needs for the test
in the display (section 6.3).
Place weight whose value flashes in the top line of the display in the center of the
weighing pan. The balance is tested.
When zero flashes, remove weight from weighing pan.
When the test is complete, the difference between the actual value and the preset
target value appears in the top line marked by "" and "D". The balance then
returns automatically to the weighing mode.
A tare value displayed before the test reappears in the display.
If a printer is attached, the test result is printed out automatically.
You can terminate the test at any time by pressing the N key.
The deviation determined by the balance depends on the ambient conditions and the setting of the balance
parameters under "Weight" in the menu, particularly on the setting of the repeatability "Repro"!
If immediately following a calibration (adjustment) with the internal weight the difference to your weight
standard is greater than that specified under "long-term stability" in the technical data, you can configure
your balance to your weight (see Appendix, section 8.4).
Flashing of the status display "AutoCal" or "Cal" signals that the balance should be calibrated and is
stopped by the check.