Mettler Toledo PR/SR Operations Instructions Page 86

  • Download
  • Add to my manuals
  • Print
  • Page
    / 124
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 85
84
PR/SR balancesMenu – Balance settings – " SYSTEM "
Specifying header contents – " Header "
Press SmartBar below "Header".
The following information can be incorporated in the header:
"Date" Current date
"Time" Current time
"BalID" Balance identification (printed out together with the balance type
and serial number)
"LotID" Identification of the sample series. The actual identification is
entered in the weighing mode.
"SampleID" Identification of the individual samples in weighing (not in header
printout). The actual identification is entered in the weighing
mode.
Mark desired information with SmartBar or cancel marking of information not
wanted.
Confirm settings with H.
In contrast to the entry of the date, time, etc. (see section 6.4.4), here "Header" appears in the top line of the
display.
If the lot and/or sample identification is activated, you are automatically prompted for their entry before
every printout, e.g. "LID= …".
Header WeightTransfer
Printout
Manual Auto AutoNoZeros
WgtTransfer
Select transfer mode – " WeightTransfer "
Press SmartBar under "WeightTransfer".
Select desired setting.
"Manual" The balance transfers the next stable weighing result to the
recording device, e.g. a printer only after the G key has
been pressed (factory setting). In weighing applications the
final results of a weighing series are automatically transferred
at the end of the series.
"Auto" The balance automatically transfers all stable weighing
values after taring incl. zero values as well as all results
which have been received for further calculation in the
weighing application (see "WeightEntry" in the applications,
section 6.1).
"AutoNoZeros" Same transfer mode as "Auto", but without displayed values
less than 30 display increments (factory setting).
Confirm settings with H.
Press H key again to return to the selection under "System".
Cal / Test
0
Menu
Settings Printout Beep
System
Cal / Test
0
Menu
Cal / Test
0
Menu
SampleID
Header
Date Time BalID LotID
Header