Mettler Toledo PR/SR Operations Instructions Page 83

  • Download
  • Add to my manuals
  • Print
  • Page
    / 124
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 82
81
PR/SR balances
Cal / Test
0
Menu
proFact CalInt VariCal
Calibration
ExtWgt InfoOn InfoOff
VariCal
2000 3000 4000 5000
ExtWeights
History
Calibration
05.01.97 07:30 Next
4 CalInt
09:20:25
- CALIBRATION HISTORY –-
05.01.96
PR5002DR
1114051374
Your entry
METTLER TOLEDO
Type:
SNR:
Bal:
1 Int 03.01.97 07:30
2 Int 04.01.97 07:30
3 Ext 04.01.97 10:15
CalExtW: 5000.00 g
4 Int 05.01.97 07:30
With G all recorded adjustment operations can be printed out if a printer is
attached.
Menu – Calibration and test function – "␣ CAL␣ "
"VariCal"
In this setting you can use your own weights for the adjustment. The balance
uses the flashing status symbol "Cal" (see section 1.5) to signal that it should be
adjusted.
"InfoOn" Display of the status message "Cal".
"InfoOff" Status message "Cal" switched off.
"ExtWeights" Selection of an external weight value to adjust the balance in
the vicinity of the usual load.
With certified balances, the certification regulations of your country may possibly
preclude use of this setting.
Section 2.5 describes how you adjust the balance with an external weight.
"History"
Recording the last adjustment operation with time and date. In this case, the type
of adjustment (internal/external) and, if applicable, the weight used are
specified. With "Next" you can recall the last 50 adjustment operations.
.
Page view 82
1 2 ... 78 79 80 81 82 83 84 85 86 87 88 ... 123 124

Comments to this Manuals

No comments