Applications – Plus/minus statistics
a) Setting known nominal weight using numeric keys
➜ Press SmartBar under "Nom", "Nom=..." is displayed in the top line.
➜ Enter numeric value of nominal weight in weight unit 1 "Unit1".
Possible values for "Nom": 0 ... max load. An entry outside the admissible range
leads to the message "Illegal value".
➜ Confirm entry with H.
This automatically sets the tolerance limits to ±2.5 % of the nominal weight first
entered. If you wish to retain these limits, press H again.
➜ Press SmartBar under "SetRef".
1. Setting reference values
This entry is necessary only in the weight check with respect to a target value or in
the weighing of equal amounts of samples using the METTLER DeltaTrac.
➜ Press SmartBar under "/".
b) Setting nominal weight using the "Set100%" function
This function is available only if the parameter "+/–%" is actvated (marked) in the
menu under "+/–Stats".
➜ Place reference (nominal) weight on weighing pan.
➜ Press SmartBar under "Set100%". The display shows 100% after the stability
was reached.
The tolerance limits +Tol and -Tol are automatically set to ±2.5% of the nominal
weight. If you wish to retain these limits, press H.