Mettler Toledo PR/SR Operations Instructions Page 50

  • Download
  • Add to my manuals
  • Print
  • Page
    / 124
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 49
48
PR/SR balances
Application
Nom -Tol +Tol Set100%
823.60 g
AutoCal BPTGNet
Cal / Test
0
Menu
0.00 g
Application
M+ n=0 x=0 Diff
0.00 g-
Application
SetRef Results Clear
0.00 g-
Nom -Tol +Tol Set100%
0.00 g
1 2 3 4 5 6 7 8 9 0 . <
Nom=50_ g-
AutoCal BPTGNet
AutoCal BPTGNet
AutoCal BPTGNet
AutoCal BPTGNet
0.00 g-
Cal / Test
0
Menu
Applications – Plus/minus statistics
a) Setting known nominal weight using numeric keys
Press SmartBar under "Nom", "Nom=..." is displayed in the top line.
Enter numeric value of nominal weight in weight unit 1 "Unit1".
Possible values for "Nom": 0 ... max load. An entry outside the admissible range
leads to the message "Illegal value".
Confirm entry with H.
This automatically sets the tolerance limits to ±2.5 % of the nominal weight first
entered. If you wish to retain these limits, press H again.
Press SmartBar under "SetRef".
1. Setting reference values
This entry is necessary only in the weight check with respect to a target value or in
the weighing of equal amounts of samples using the METTLER DeltaTrac.
Press SmartBar under "/".
b) Setting nominal weight using the "Set100%" function
This function is available only if the parameter "+/–%" is actvated (marked) in the
menu under "+/–Stats".
Place reference (nominal) weight on weighing pan.
Press SmartBar under "Set100%". The display shows 100% after the stability
was reached.
The tolerance limits +Tol and -Tol are automatically set to ±2.5% of the nominal
weight. If you wish to retain these limits, press H.
Page view 49
1 2 ... 45 46 47 48 49 50 51 52 53 54 55 ... 123 124

Comments to this Manuals

No comments