PR/SR balancesApplications – Formulation
Entering target weight
Press SmartBar under "Weight", "Trg=…" appears in the top line of the display.
Confirm entry with H.
Press H again, the balance returns to the weight display.
Enter numeric value of target weight in weight unit 1 ("Unit1").
Possible values for "Trg": 0 … max. load. An entry outside the admissible range
leads to the message "Illegal value".
The METTLER DeltaTrac changes to the +/– display with pointers. Both pointers
point to 9 o’clock. If "CmpDiff" has been selected under "Results", the top line
switches to the display of the difference to the target weight, i.e. the preset target
value appears with a negative sign, marked by a "".
Weigh in first component of the formula.
When the weight approaches the target weight, the display approaches zero and
the pointers of the DeltaTrac reach the 6 o'clock position. The tolerance markers
correspond to a deviation of ±2.5 % from the target weight.
Press SmartBar under "M+".
The next stable weighing result is transferred to the net memory, the display is
automatically set to zero and the item counter "n=…" incremented.
At the same time, the preset value of the target weight is cleared and the DeltaTrac
automatically returns to the display of the remaining range.
0.00 g
0.00 g
Weight SampleID
1 2 3 4 5 6 7 8 9 0 <
Trg=5.00_ g
NextComp
-5.00 g-
M+ n=0 Results Comp.
Cal / Test
0
Menu
AutoCal
Cal / Test
0
Menu
M+ n=0 Results Comp.
-0.07 g-
M+ n=1 Results Comp.
0.00 g-
AutoCal
BPTGNet
AutoCal
BPTGNet