PR/SR balancesStartup procedure
Sequence of the calibration (adjustment) of the balance with an internal
weight
Requirement
The calibration setting in the menu corresponds to the factory setting "proFACT" (see
section 6.3).
If the balance wishes to calibrate (adjust) itself, "AutoCal" flashes in the display.
On completion of the current weighing series or when the work allows, remove
any load from weighing pan.
After a few minutes, the calibration (adjustment) is triggered automatically by the
balance.
Or
initiate the calibration (adjustment) manually by pressing and holding the
key and by pressing the SmartBar under the word "Calibration".
During the calibration, "BALANCE CALIBRATION" appears in the bottom line of the
display.
Finally, a completed calibration (adjustment) is confirmed by "Cal done" and the
balance returns automatically to the weighing mode. A tare value displayed
before the calibration reappears in the display.
If a printer is attached, the adjustment can be recorded automatically, see
section␣ 6.3.
M+ n=0 =0.00g M-
345.78 g-
Test Calibration Repro
Start
Cal done
- BALANCE CALIBRATION
AutoCal
10:57:59
METTLER TOLEDO
Type:
SNR:
Bal:
Int. calibration done
Signature:
........................
PR5002DR
1114051374
Your entry
– BALANCE CALIBRATION ––
––––––––– END –––––––––-
17.06.95
Cal / Test
0
Menu
You can terminate the calibration procedure at any time by pressing the N key.
"Signature" appears on the printout if the calibration was started using the keypad.
Within a weighing series (item counter n > 0), the calibration is not initiated automatically, but it can be
started using the key.