34
PR/SR balances
Applications – Dynamic weighing
➜ Add weighing sample to container.
➜ Press SmartBar under "Start", the weighing then starts immediately.
During the weighing, a "countdown" runs in the display and the e symbol
flashes.
➜ After elapse of the weighing time, the weighing result appears in the bottom line
of the display as "dw=...". The item counter "n=..." is incremented by one.
"dw" and "n" remain in the display until the values are overwritten by a new result
or cleared with "Clear".
• You can terminate the current weighing at any time with the N key or restart it with "Start". If the balance is
overloaded during the weighing, the weighing is automatically aborted and the message "Abort" outputted.
• Depending on the selection under "Results", you can also display values other than "dw" in the bottom line.
• If you have activated the sample identification in the menu (section 4.3), you are prompted for its entry with
"SID=…" before every start.
• You can also initiate the weighing using the LC-FS foot switch (Optional equipment, section 7.3). The foot
switch must be set to switch position 4 j for this. You thus have both hands free and need not touch
the balance.
Dynamic weighing with manual start
➜ Place empty container on balance and tare.
AutoCal BPTGNet
0.00 g-
Start n=0 dw=0.00g
AutoCal BPTGNet
423.45 g-
Start n=0 dw=0.00g
AutoCal BPTGNet
--2--
Start n=0 dw=0.00g
AutoCal BPTGNet
--1--
Start n=0 dw=0.00g
AutoCal BPTGNet
420.12 g-
Start n=1 dw=420.12g
BPTGNet
BPTGNet
BPTGNet
2. Weighing
Depending on the setting in the menu (section 6.1.2)
• the weighing must be started either manually (factory setting),
• or automatically when a load is placed on the balance. The first weighing must
be started manually in this case.