Mettler Toledo PR/SR Operations Instructions Page 78

  • Download
  • Add to my manuals
  • Print
  • Page
    / 124
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 77
76
PR/SR balances
1 2 3 4 5 6 7 8 9 0 <
F=1.3_
Unit Unit2 CustomUnit
Weighing
Off
On Define
CustomUnit
1 2 3 4 5 6 7 8 9 0 <
DP=_
A-Z a-z 0-9*! <
Name=_
Define
1
/x
Cal / Test
0
Menu
Cal / Test
0
Menu
Cal / Test
0
Menu
Cal / Test
0
Menu
6.2.5 Freely definable unit - "CustomtUnit"
If you activate the unit you have defined, you can switch between this and the other units in every application. The displayed
value in "CustomUnit" corresponds to the inputted factor "F" multiplied or divided by the weight determined by the balance in
grams.
Menu – Weighing parameters – " WEIGH "
Under "Weighing", select the freely definable unit "CustomUnit".
The following selection appears in the bottom line of the display:
"Off" CustomUnit not activated (factory setting)
"On" CustomUnit available in weighing. Results appear followed by
"[C]" or with user-defined name
"Define" Define conversion factor, display format, name of "CustomUnit" or
factor divided by weight "
1
/x" function.
Specify the unit under "Define".
Unit1 Unit2 CustomUnit
Weighing
Factor DecPlaces Name
Define
Off On Define
CustomUnit
Under "Factor", enter numeric value of the multiplication factor "F=…" (-sign first
if factor negative), close entry with H.
Possible values for "F": ±0 to 100 000, factory setting: F=1.
If an entry is outside the admissible range, the message "Illegal value" appears.
Under "DecPlaces", enter the number of decimal places "DP=…" which should
be displayed, close entry with H.
Possible values for "DP": 0 to 7, factory setting: DP=2.
If an entry is outside the admissible range, the message "Illegal value" appears.
Press SmartBar under "On" to activate the "CustomUnit" you have defined.
Confirm settings with H and return to previous selection.
Enter alphanumeric characters of your designation of the custom unit "Name = "
under "Name", confirm entry with H.
Max. 4 characters are possible. Names of weight units, e.g. "g" or "oz" are not
admissible. Factory setting: Name = [C].
If entries are outside the admissible range, the message "Illegal value" appears.
Press the SmartBar under
1
/x to activate this function, if you want to use your unit
for conversion "factor divided by weight".
Confirm definition with H and return to selection "Off On Define".
Page view 77
1 2 ... 73 74 75 76 77 78 79 80 81 82 83 ... 123 124

Comments to this Manuals

No comments